FIB Preliminary

1. Swipe in to the computer logbook to activate the microscope.


IP1 and IP2 lights.

2. Ensure IP1 and IP2 green lights are on (front panel below the chamber).


DP, Water, Air Pres., DP Power

3. Make sure the green lights are on for DP, Water, and Air Press. Make sure the DP Power Switch is on.


SC Vacuum, SEC Vacuum

4. Check that both S.C. and S.E.C. Vacuum green lights are on at HIGH vacuum.


SC Air Lock Valve

5. Check that the S.C. AIR LOCK VALVE is in the closed position.


NOTE ON SWITCHES: All of the switches on the FIB are protected against accidental changes in position. In order to flip the switches, you MUST pull each switch lever OUT before attempting to change the position! Otherwise, the switch will be damaged.

NOTE ON SPECIMENS: The FIB will accommodate specimens that are 5x5 mm or 5x10 mm. Wafer thick specimens are best. No wet or contaminated specimens are allowed.

If any of the above conditions are not met or an alarm is sounding, contact ACMAL staff immediately.

Operating Procedure: Start Up